Rapid Distillation Process Analyzer
Benefits
- Cycle time 10 to 15 minutes (for diesel 10 min. and for gasoline 15 min.)
- Results within ASTM D86 requirements
- Measuring points from IBP to FBP, free programmable
- Core components meet ASTM D86 design
- Optical level measurement for dosing unit and receiver
Marking
ATEX: II 2G Ex h IIC T4 Gb X
IECEx: on request
NEC 500: Class I, Division 2, Group B,C and D
NEC 505: Class I, Zone 1, AEx d e ib px IIB resp.
IIB+H2 T3 resp.T4
CEC Sec. 18: Ex d e ib px IIC T3 resp.T4
TR CU: on request
Ambient Temperature
T4Description
To remain competitive, today’s refiners must employ all optimization and product control techniques available. The use of online physical property analyzers is one of the key features to reach those objectives because they measure important quality properties in the process directly.
Distillation is a physical method of separating the component substances from a liquid mixture by selective vaporization and re-condensation. Distillation is based on differences in volatilities of the components of the liquid mixture. The distillation curve is one of the most common quality parameters of liquid hydrocarbons such as naphtha, gasoline, kerosene, diesel and gas oil.
Technical Data
Technology | distillation |
---|---|
Method | correlates with: ASTM D86, DIN EN ISO 3405, IP 123 |
Sample at inlet | filtered 50 μm, no suspended water, bubble-free, transparent, Cloud Point 0 °C 20 to 40 l/h 1 to 3 bar (14.5 to 43 psi) max. 50 °C, temperature change: max. 1°C/min, |
Sample at outlet | atmospheric pressure |
Electrical data | max. 8 outputs (4 to 20 mA; 1000 Ω), active isolated on request 4 to 20 mA; 160 Ω DC 24 V; max. 0.5 A; sum alarm Ready/Come-Read, Power identification Validation identification, Analysis Cycle Active high: DC 15 to 28 V; low: DC 0 to 4 V Reset, Inhibit, Stream request, Validation request, Decoking request, Automatic stream switching |
Communication | MODBUS RTU/TCP (RS485, RS422, VDSL/FO (IS) MODBUS/TCP via FOC is remote software with modem, ISDN, Ethernet via VDSL modem FO (FS) |
Ambient conditions | operation 5 to 40 °C (41 to 104 °F) storage -20 to 60 °C (-4 to 140 °F) operation 5 to 80 % relative humidity, non-corrosive; storage 5 to 80 % relative humidity, non-corrosive IP 54 (comparable with NEMA 13) |
HMI | touch display 19" remote access on request cyber security operating system Windows10 enterprise LTSB Bartec Benke control software PACS |
Process connections | tube fittings: typically Swagelok 6mm/12mm/18mm; others on request sample inlet up to 2 sample streams and 1 validation stream sample outlet vent line (if required) coolant (if required) |
Weight and dimensions | approx. 300 kg (W x H x D) approx. 1150 x 1900 x 710 mm free space right: 150 mm/left: 100 mm |
Application Notes
-
BENKE Rapid Distillation Process Analyzer rapiDist-4 Atmospheric Distillation Process Control with Advanced Distillation Process Analyzer Technology
- EN
Application Note; VISC-4; VI-4; Lube Oil
Questionnaires/Requirement Sheets
-
Questionnaire_PL AMS_PAT
Data Sheets
-
Rapid Distillation Process Analyzer rapiDist-4